Maquet Servo-AIR: touchscreen not responding
What happened
On a Maquet Servo-AIR ventilator the touchscreen stopped working. Before ordering a replacement, the engineer had to confirm whether the touch layer is included in spare part 6882638, since a display without touch had been encountered under that same part number before.
What was checked
- Checked the scope of spare part 6882638 as reported in the case — whether the touch layer ships with the display.
Cause
The touch layer of the display module had failed. The practical obstacle was the ambiguity around the part number, which the engineers resolved by confirming its actual scope.
What was done
As reported in the case, part 6882638 was confirmed to be supplied as 'Display incl. touch screen', so the display and the touch layer are replaced together as a single part.
Safety
Before opening the display assembly, switch the ventilator off and disconnect it from the mains, and before it returns to patient use run the electrical safety test and a ventilation check on a test lung.
Lesson: Confirm the exact scope of a display part number before ordering — 'display' and 'display with touch screen' are not always the same supply, and a wrong assumption means a second visit.
Covered in the course: Ventilator Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.