Maquet SERVO-i: Oxygen sensor and expiratory cassette hour counters need a reset
What happened
The engineers needed to clear the operating-hour counters for the oxygen sensor and the expiratory cassette on a SERVO-series ventilator and asked how this is done correctly.
What was checked
- Confirm that the oxygen cell and the cassette membrane have actually been replaced or are genuinely due — a counter reset must never stand in for a worn consumable.
- After the reset, check that the hour counters restart from zero and that the relevant system screens show the consumables as new.
Cause
The hour counters are usage trackers for replaceable consumables — the oxygen cell and the expiratory cassette membrane. When they reach their limit the ventilator prompts replacement, and after renewal the counters must be cleared. In this case the engineers found that the cassette allows only the membrane hours to be reset locally, while the full reset on the SERVO-i is deliberately restricted to authorised service, so there is no legitimate way to clear both counters from the user level.
What was done
Only the membrane hours can be reset on the spot on the expiratory cassette. The full reset of the oxygen sensor and cassette hour counters on the SERVO-i is a procedure from the manufacturer's service documentation, done by authorised personnel. The engineers reported the case solved after that reset was carried out; no counters were cleared on worn parts and no workarounds were used.
Safety
Before opening the expiratory cassette or handling the oxygen sensor, disconnect the ventilator from the patient and from mains power; after any counter reset, verify the oxygen measurement against a calibrated gas source and run the full manufacturer's functional check before return to clinical use.
Lesson: Hour counters exist to force genuine renewal of consumables, so reset them only after the part has actually been replaced and only through the manufacturer's documented procedure.
Covered in the course: Ventilator Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.