Medica EasyBloodGas: blocked sampling channel
What happened
Suspected obstruction in the sampler / fluidic path. The case shared the clearing procedure rather than the exact presenting symptom.
What was checked
- Pull-through with a syringe connected to the sampler needle: draw the plunger back to create vacuum and pull fluid through the channel, as reported in the case.
- Disconnect the pump tube and push air through it with a syringe.
- Repeat the daily cleaning cycle.
- Run a calibration check afterwards to confirm the analyzer is fit for use.
Cause
Not stated in the case. The engineers attributed the fault to sample residue partially blocking the fluidic channel — a probable cause, not a confirmed one.
The cause is probable and was not fully confirmed.
What was done
The fix in this single case was mechanical, no parts involved: a syringe pull-through through the sampler needle, then an air purge through the pump tube, followed by daily cleaning and a calibration check. The outcome was reported as probable — treat results as unverified until calibration passes.
Safety
Treat all fluid in the sample path as potentially infectious: wear gloves and eye protection, and finish cleaning plus a calibration check before the analyzer returns to patient samples.
Lesson: Before blaming the pump or electronics, clear a sluggish fluidic path with a syringe pull-through and an air purge, then confirm with calibration.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.