Medica · EasyBloodGas

Medica EasyBloodGas: blocked sampling channel

What happened

Suspected obstruction in the sampler / fluidic path. The case shared the clearing procedure rather than the exact presenting symptom.

What was checked

Cause

Not stated in the case. The engineers attributed the fault to sample residue partially blocking the fluidic channel — a probable cause, not a confirmed one.

The cause is probable and was not fully confirmed.

What was done

The fix in this single case was mechanical, no parts involved: a syringe pull-through through the sampler needle, then an air purge through the pump tube, followed by daily cleaning and a calibration check. The outcome was reported as probable — treat results as unverified until calibration passes.

Safety

Treat all fluid in the sample path as potentially infectious: wear gloves and eye protection, and finish cleaning plus a calibration check before the analyzer returns to patient samples.

Lesson: Before blaming the pump or electronics, clear a sluggish fluidic path with a syringe pull-through and an air purge, then confirm with calibration.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.