Medizs CP-11: projector unresponsive to all controls after a power outage
What happened
After a mains power interruption the medical projector stopped responding to every button except the lamp on/off control. No error codes were reported in the case.
What was checked
- Verified that the unit was powered and that only the lamp on/off control worked
- Contacted the manufacturer's service support and described the exact behaviour
Cause
The engineers treated it as a corrupted internal settings/state after the sudden power loss; the exact cause was not determined. This is one reported case, not a rule.
What was done
A reset was performed according to a procedure from the manufacturer's service documentation, guided by the manufacturer's service support and done by authorised personnel; the button sequence itself is not reproduced here. Normal response to all controls was restored after the reset.
Safety
After any power-related fault, confirm the device completes its full start-up self-test and passes the electrical safety checks required before it is returned to clinical use.
Lesson: A sudden power cut can leave a device in a stuck internal state, so check the manufacturer's documented reset procedure with service support before suspecting a hardware fault.