Medonic M series: constant background added to WBC and PLT counts
What happened
A stable background added directly to WBC and PLT counts, with a DE flag on WBC, while RBC measured normally with no blood flags — pointing to the WBC/PLT counting path rather than the whole fluidics.
What was checked
- Hydraulic system cleaned with bleach and cleanliness afterwards confirmed, so internal line contamination was largely excluded.
- Reagents cross-tested on another analyzer with no background there, so reagents were ruled out.
- Tubing at the pinch valves and the rotary valve gland seals replaced; no leaks or salt deposits found.
- Inspection for air ingress, first of all at the connection between the WBC/PLT chamber and the aperture.
- Two-head drain pump checked for brush wear and sparking; the waste pump was relocated as a test.
- Chamber electrodes and their soldered joints inspected for oxidation, and the power socket and grounding checked.
- Readings compared with another analyzer, including swapping instruments for comparison.
Cause
Not finally established in this case. The most likely contributors reported in the case were air ingress at the chamber-to-aperture connection, non-original tubing on the pinch valves, and brush wear or electrical noise from the drain pump — all of which would affect the WBC/PLT chamber while RBC stays clean.
The cause is probable and was not fully confirmed.
What was done
Probable, not fully confirmed on this unit. In similar reported cases the background was cleared by replacing the hydraulic tubing with original manufacturer tubing instead of cheap aftermarket ones, and in another case by relocating the waste pump; the engineers were advised to check the chamber-to-aperture connection for air ingress and the condition of the drain pump.
Safety
Before opening the fluidics compartment or working on pumps and valves, switch off and isolate the analyzer from the mains and relieve line pressure, treat all fluid and waste as potentially infectious biohazard, and wear eye and skin protection when handling bleach.
Lesson: When a stable background persists on WBC/PLT after cleaning and reagent checks, suspect the counting-path hardware — original tubing quality, air-tightness of the chamber-to-aperture connection and drain pump condition — before blaming the electronics.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.