Medonica Hem: unstable results from a pinhole in pinch-valve tubing
What happened
A hematology analyzer was reported with unstable results and no error codes. In this case, inspection found a small hole in the silicone tube of the third pinch valve.
What was checked
- With unstable results and no error codes, inspect the silicone tubing at every pinch valve along its full length; in this case the hole was in the tube of the third valve.
- Gently flex the tube near the pinch point — a pinhole often becomes visible only when the tube moves.
- Look for signs of air ingress: bubbles in the fluid line and a raised or drifting background during aspiration, as reported in the case.
- Before ordering, confirm size and article; reported in the case: silicone tube 1.5 × 3.0 mm, part 9970002 (3 m), available from official suppliers.
Cause
A pinhole defect in the silicone tube let the system draw in air at the third pinch valve; the air gave a raised background and dosing errors, which the engineers linked to the unstable results.
The cause is probable and was not fully confirmed.
What was done
Replace the affected tube with new silicone tubing of the same size — reported in the case: 1.5 × 3.0 mm, part 9970002 (3 m), from official suppliers. Then flush the fluid path and recheck background and dosing with control material; the engineers considered this fix probable, so confirm stability before returning the unit to routine use.
Safety
Treat the removed tube and any fluid in it as potentially infectious waste: work in gloves and eye protection, and disinfect and dispose of it according to local biohazard rules.
Lesson: When a hematology analyzer drifts for no clear reason, check wear-prone tubing at the pinch valves first — a pinhole can imitate calibration or reagent problems.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.