Medonica M20: worn rotary valve seals
What happened
Worn rotary valve seals on a lab analyzer: the large seal was easy to order as a spare, but the small one was hard to find.
What was checked
- Since the original small seal was hard to source, the engineers measured it and searched for an analogue by its dimensions, as reported in the case.
- The matched NBR seal was compared with the original and test-fitted: it matched by dimensions but needed additional sealing with a washer to seat properly.
Cause
The rotary valve seals had simply worn out in service; the case reported no other cause. The large seal is a routinely available spare, while the small one is not easy to obtain.
What was done
The large seal was ordered as a regular spare. For the small one, the engineers found a dimensional analogue — an NBR hydraulic seal (K21-003/5, 3x8x4.5x5x2.5 as reported in the case). Despite matching the original by dimensions, it needed an extra washer to seal properly; with the washer fitted, the valve sealed and the fault was resolved.
Safety
Before opening the fluid path, power down and isolate the analyzer, treat any fluid that has contacted samples as potentially infectious, and complete leak and quality-control checks before returning the analyzer to use.
Lesson: For a hard-to-source small seal, a dimensional analogue from a general hydraulic seal catalogue can be a workable substitute, but expect to adjust the fit and verify sealing before the unit goes back into service.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.