MedRad Stellant: syringe not recognized on the first try
What happened
The injector did not always recognize the syringe when it was seated, so users had to remove it and re-insert it before the system accepted it.
What was checked
- Checked the ribbon cable to the sensors under the cover on the syringe side.
- Checked the flex rings (magnetic rings) and the Hall sensors that read them.
- Checked the optical syringe sensors at the ends of the syringe holder.
Cause
The engineers found contrast media residue on the syringe detection sensors, with probable corrosion of the contacts on a Hall sensor; in this case that was named as the probable cause of the intermittent recognition.
The cause is probable and was not fully confirmed.
What was done
Following the manufacturer's service documentation, the engineers removed the main and interface boards, released the retaining nuts and removed the syringe holder (eight screws, as reported in the case), washed the flex rings, Hall sensors and optical sensors clean of contrast residue, re-soldered the Hall sensor contacts where corrosion was found, and reassembled in reverse order. The outcome in this case was reported as probable: the work matched the symptoms, but a lasting result was not confirmed.
Safety
Unplug the injector from mains before opening the front end, treat contrast residue as a biological hazard and work in gloves and eye protection, and complete electrical safety checks before returning the injector to clinical use.
Lesson: Intermittent syringe recognition usually points to contaminated or corroded sensors in the holder, so clean and inspect them before suspecting the electronics.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.