МЕК (Medica/Mindray MEK) · MEK 6400

MEK 6400: bubble fault in volumetric tube with no visible leak

What happened

The analyzer repeatedly reported the BBL1 bubble fault in the volumetric tube during measurements, while no fluid leak was visible anywhere in the system.

What was checked

Cause

The engineers found that air was being sucked in under a rubber cuff (seal) in the volumetric tubing — an air ingress that does not show up as a fluid leak.

What was done

Replacing the rubber cuffs (seals) of the volumetric tube stopped the air ingress, and the bubble fault no longer appeared on repeated measurements.

Safety

Treat all fluid-path parts, spills and waste from a hematology analyzer as potentially infectious: wear gloves and eye protection when opening the tubing, and disinfect any spilled diluent or sample.

Lesson: A bubble error without a visible leak usually means air is being drawn in at a seal on the suction side, so check the rubber cuffs and joints before suspecting electronics or the counting chamber.