Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Melag 23 B+
Melag · 23 B+

Melag 23 B+: no water feed error after warm-up

What happened

The autoclave reported a no water feed error after warm-up. When the water feed pump was taken apart, it ran for a couple of cycles, as reported in the case, and then the error came back.

What was checked

Cause

The pump itself was faulty and jammed unpredictably, so cleaning and reassembly did not address the root cause in this case.

What was done

The pump was replaced with a new one, and the fault was resolved.

Safety

Before opening a steam sterilizer, isolate it from the mains, let the chamber and pipework cool down, and confirm the pressure is fully released — residual hot water and steam can cause severe burns.

Lesson: If a pump works only briefly after being taken apart, it is worn, not repaired — once filters and water supply are ruled out, replace it instead of chasing an intermittent fault.

Covered in the course: Steriliser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes