Melag 23 B+: no water feed error after warm-up
What happened
The autoclave reported a no water feed error after warm-up. When the water feed pump was taken apart, it ran for a couple of cycles, as reported in the case, and then the error came back.
What was checked
- The engineers disassembled and inspected the water feed pump; it pumped for a couple of cycles, then failed again.
- Mesh filters at the vacuum pump inlet and outlet were proposed for cleaning to rule out a flow restriction in the water path.
Cause
The pump itself was faulty and jammed unpredictably, so cleaning and reassembly did not address the root cause in this case.
What was done
The pump was replaced with a new one, and the fault was resolved.
Safety
Before opening a steam sterilizer, isolate it from the mains, let the chamber and pipework cool down, and confirm the pressure is fully released — residual hot water and steam can cause severe burns.
Lesson: If a pump works only briefly after being taken apart, it is worn, not repaired — once filters and water supply are ruled out, replace it instead of chasing an intermittent fault.