Melag 23 B: sterilizer fault from clogged vacuum pump inlet strainer
What happened
The exact fault was not specified in the case: engineers asked what to look at on a Melag 23 B autoclave that had failed. The discussion focused on the vacuum side.
What was checked
- Inspect the filter inside the chamber.
- Inspect the strainer at the vacuum pump inlet.
- Check that the door gasket is seated correctly in its groove.
- Check all filters, including the two lower screen filters.
Cause
In this one case, the strainer at the vacuum pump inlet was clogged with contamination described as biological in character, although, as reported in the case, it had been cleaned about three months earlier.
What was done
The engineers cleaned or replaced the clogged strainer at the vacuum pump inlet, and the fault was reported as solved; the remaining filters were checked as needed.
Safety
Before opening the chamber or removing any filter, confirm the autoclave is fully depressurized and cooled and isolated from the mains, and handle the contaminated strainer and filters as biohazard material, with gloves and disinfection.
Lesson: A vacuum sterilizer can be choked by a pump intake strainer that has fouled again within months of cleaning, so check it early and look for the reason it clogs so fast.
Covered in the course: Steriliser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.