MELAG Vacuklav 23B+: vacuum pump starts slow at atmospheric pressure

What happened

At the start of the sterilization program, while the chamber was still at atmospheric pressure, the vacuum pump ran at very low speed (reported in the case, up to about a minute); once a minimal vacuum was reached, the speed came up smoothly and the pump then ran normally. No error codes appeared, and the vacuum test passed.

What was checked

Cause

In this case the pump had been put back together incorrectly during an earlier repair: rubber diaphragms and valves were fitted in each other's places. At atmospheric pressure this caused slipping and low speed, yet the pump could still build the required vacuum afterwards — which is why routine checks and the vacuum test showed nothing.

What was done

The engineers rebuilt the pump and installed the diaphragms and valves in their correct positions; the slow start disappeared and the pump worked normally. No parts were replaced. The rebuild was done to the manufacturer's service documentation.

Safety

Before opening the pump, isolate the sterilizer from the mains, let it cool and confirm that the chamber and tubing are depressurized; after any pump work, run the vacuum test and a full cycle and verify the sterilization result before returning the unit to use.

Lesson: A fault that shows up only in one phase of a cycle (here only at atmospheric pressure, at the very start of evacuation) points to the internal condition of the mechanism — including how it was left after the last rebuild — before you chase connections, filters and settings.

Covered in the course: Steriliser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.