Merivaara · Optima

Merivaara Optima: table beeps on any key press, actuators do not move

What happened

The operating table gives a beep on every press of the hand control, but no section actuator moves. No error codes were shown in the reported case.

What was checked

Cause

In the reported case, the cause was a power cable with poor contact to the table; the fault state itself stays latched until it is cleared by the documented reset, so the beeping can persist even after the physical cause is gone.

What was done

The engineers replaced the defective power cable and cleared the fault indication with a procedure from the manufacturer's service documentation, done by authorised personnel. In the reported case the table then moved normally on all sections and the symptom did not return.

Safety

Isolate the table from the mains and confirm zero voltage before handling the power cable or its connector, and run an electrical safety test before the table returns to clinical use.

Lesson: A fault beeper with all drives blocked is often a mundane supply-side problem, so check the power path first and clear fault states only through the documented procedure.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.