Micro CC-20: analyzer stops printing results
What happened
The analyzer ran the test but did not print the results report.
What was checked
- As reported in the case, a full software reload onto the CF card was considered first, since the fault looked like a system problem.
- The engineers checked the results database file TDL.dat and found it damaged.
Cause
In this case the analyzer would not print results because the TDL.dat database file was corrupted.
What was done
The engineers replaced the damaged TDL.dat file with a copy supplied by the manufacturer's authorised representatives; printing returned to normal and a full software reload was not needed.
Safety
Isolate the analyzer from the mains before removing or replacing internal storage, and complete the electrical safety checks before returning the unit to use.
Lesson: Before planning a full software reinstall, check the integrity of the instrument's database files, because a single corrupted file can stop result printing.