Micro · CC-20 Plus

Micro CC-20 Plus: diluent level sensor operating principle

What happened

A field question about the diluent level sensor: a plastic rod spanning the full height of the container, two wires, no float, and a reported closure at the rod tip when liquid is present.

What was checked

Cause

No fault in this case. The engineers established the sensing principle: the sensor is conductive (conductometric) — the tip and the second conductor form a small circuit that closes through the salts dissolved in the water/diluent.

What was done

No repair was needed: it was a working question resolved by explanation. The engineers agreed the rod is a conductive level sensor; nothing was replaced or adjusted.

Safety

Before inspecting any level sensor or fluid path, isolate the analyzer from mains power and handle the container and any spilled diluent or reagent as potentially biohazardous.

Lesson: Knowing the sensing principle of a component — conductivity through the liquid, in this case — often explains the observed behaviour without any part replacement.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →