Micro CC-20 Plus: diluent level sensor operating principle
What happened
A field question about the diluent level sensor: a plastic rod spanning the full height of the container, two wires, no float, and a reported closure at the rod tip when liquid is present.
What was checked
- As reported in the case: the probe is a plain plastic rod with two wires and no float or moving parts; one of the two is the cable shield.
- One suggestion in the discussion was that the volume is only calculated in software; the physical rod with a closure at its tip ruled that out.
- The engineers considered where the electrical path actually closes and found it closes at the rod tip once the liquid reaches it.
Cause
No fault in this case. The engineers established the sensing principle: the sensor is conductive (conductometric) — the tip and the second conductor form a small circuit that closes through the salts dissolved in the water/diluent.
What was done
No repair was needed: it was a working question resolved by explanation. The engineers agreed the rod is a conductive level sensor; nothing was replaced or adjusted.
Safety
Before inspecting any level sensor or fluid path, isolate the analyzer from mains power and handle the container and any spilled diluent or reagent as potentially biohazardous.
Lesson: Knowing the sensing principle of a component — conductivity through the liquid, in this case — often explains the observed behaviour without any part replacement.