Micros ES60: dispensing syringe seal grease replacement
What happened
Lubrication on the dispensing syringe O-rings needed renewal, so the engineers had to select a suitable replacement for the original grease.
What was checked
- Searched for a lubricant that does not create counting background in a hematology analyzer, since grease traces can reach the sample path.
- Compared the candidate grease with the original one from the ES60 repair kit for consistency before applying it to the O-rings.
Cause
In this case, the original grease on the dispensing syringe O-rings was worn out or depleted, leaving the seals running dry.
What was done
The engineers applied a third-party background-free grease with a consistency close to the original repair-kit lubricant to the syringe O-rings; vaseline was rejected as unsuitable. The proper repair-kit grease for the ES60 was also ordered for future service, and the case was reported as solved.
Safety
Syringe seals contact blood-containing fluid, so work in gloves and confirm the analyzer passes quality control before returning it to use.
Lesson: On analyzer fluidics, use the manufacturer's repair-kit grease whenever possible; any third-party substitute must be background-free and seal-compatible, and household options like vaseline are never acceptable.