Mikro CC-20: through hole in the leukocyte counting chamber
What happened
On a hematology analyzer, a through hole was found in the wall of the leukocyte counting chamber, directly next to the aperture.
What was checked
- The engineers inspected the counting chamber and located a through opening in the wall beside the aperture.
- The damage was attributed to the aperture burn-cleaning method used earlier on this instrument; no other cause was identified in this case.
Cause
Most probable cause: the burn method for clearing a clogged aperture overheated the chamber wall and burned a hole next to the aperture. The engineers noted that this method is now rarely used precisely because of such damage.
What was done
As a stopgap the hole was sealed with epoxy resin, and the analyzer worked for some time; the engineers treated this as a temporary measure only. The proper repair was replacing the counting chamber, which finally solved the problem.
Safety
Chamber contents are potentially infectious patient material: wear gloves and eye protection when opening the fluid path, decontaminate any spilled fluid, and check the repaired area for leaks before returning the analyzer to use.
Lesson: Heat-based aperture cleaning on a plastic counting chamber can burn a hole straight through the wall — follow the manufacturer's documented cleaning procedure instead of improvised methods.