Mikross-20+: Unstable water blank readings
What happened
The analyzer gave a different water blank result almost every run, with no error codes — readings varied without any repeatable pattern.
What was checked
- The measurement path was checked first — optics, cuvette, water and reagent delivery — since a wandering blank usually points there; nothing in that chain explained the variation.
- Only after going through almost the whole analyzer did the engineers isolate the fault to the power supply, whose output proved unstable because of degraded electrolytic capacitors.
Cause
Dried-out electrolytic capacitors in the analyzer's power supply. With the supply unstable, ripple reached the measurement electronics and appeared as random blank variation. The case did not confirm whether the unit was internal or external.
What was done
The engineers replaced all electrolytic capacitors in the power supply; after that the water blank became stable and repeatable, and the analyzer was returned to service.
Safety
Capacitors in a switched-mode supply keep a dangerous charge after the mains is disconnected: isolate the analyzer, verify with a meter that the DC bus is discharged, and complete the electrical safety test before returning the unit to use.
Lesson: Random, non-repeating results with no error codes can start in the power supply — check the rails early instead of dismantling the whole analyzer first.