Mikross-20+: Unstable water blank readings

What happened

The analyzer gave a different water blank result almost every run, with no error codes — readings varied without any repeatable pattern.

What was checked

Cause

Dried-out electrolytic capacitors in the analyzer's power supply. With the supply unstable, ripple reached the measurement electronics and appeared as random blank variation. The case did not confirm whether the unit was internal or external.

What was done

The engineers replaced all electrolytic capacitors in the power supply; after that the water blank became stable and repeatable, and the analyzer was returned to service.

Safety

Capacitors in a switched-mode supply keep a dangerous charge after the mains is disconnected: isolate the analyzer, verify with a meter that the DC bus is discharged, and complete the electrical safety test before returning the unit to use.

Lesson: Random, non-repeating results with no error codes can start in the power supply — check the rails early instead of dismantling the whole analyzer first.