Mindray 5150: aperture voltage alarm not cleared by rinsing

What happened

On a Mindray 5150 hematology analyzer the RBC aperture kept flagging abnormal aperture voltage; repeated aperture rinsing, reported in the case, did not clear the message.

What was checked

Cause

The cause reported in the case was diluent underfill into the aperture bath, traced to the diluent valve not delivering the full volume — reported as valve No. 13 on the 5150. In a separate 30s case the same message came from a wet chamber.

What was done

Confirm the fill volume, then service or replace the underfilling valve (valve No. 13 in the case), make sure every cover and shield seats fully, and in the wet-chamber scenario dry the chamber completely. The fault was reported cleared after this.

Safety

Treat all fluid paths and spills as potentially infectious, and after any aperture or valve work run decontamination and quality control before the analyzer goes back to patient samples.

Lesson: An aperture voltage alarm describes the conductivity around the aperture, so prove diluent volume, cover closure and chamber dryness before suspecting the aperture itself.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.