Mindray · BC-2300

Mindray BC-2300: corroded measuring chamber wire

What happened

The wire running from the measuring chamber had corroded through.

What was checked

Cause

The wire had simply rotted: moisture and sample/reagent residue around the chamber slowly corrode the conductors, and in similar units water leaking through chamber cracks gets straight into the measuring cable.

What was done

The engineers soldered the wire back onto the chamber; as reported in the case, the analyzer has been running without trouble for three years since.

Safety

Treat the chamber and its wiring as biohazardous, since sample and reagent residue remains inside, and after any soldering pass the electrical safety test — ground continuity and leakage current — before returning the analyzer to use.

Lesson: A corroded chamber wire is a moisture problem, not just a wire problem — find and fix the leak path into the chamber and cable, or the joint will not last.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.