Mindray BC-3000 Plus: WBC chamber will not drain
What happened
Hematology analyzer: no drain from the WBC (leukocyte) chamber; the engineers were tracing which valve is responsible for draining it.
What was checked
- Trace the fluid path from the WBC chamber to identify which valve handles its drain.
- Disassemble and inspect the suspected valve for debris build-up.
Cause
The suspected cause, reported in the case, was dirt accumulating inside the drain valve, restricting flow and degrading the pumping of fluid out of the chamber.
The cause is probable and was not fully confirmed.
What was done
The engineers found the recommended fix was to disassemble and clean the valve, since dirt builds up in it. The outcome was reported as probable: the specific valve and the final result were not confirmed in the case.
Safety
Before opening the fluid path, isolate the analyzer from mains power, and treat all internal fluids and wetted surfaces as potentially infectious biohazard; wear gloves and eye protection.
Lesson: When a chamber will not drain, trace the fluid path to the responsible valve and clean it before considering replacement.