Academy / Field fault notes / Mindray BC-3000 Plus
Mindray · BC-3000 Plus

Mindray BC-3000 Plus: WBC chamber will not drain

What happened

Hematology analyzer: no drain from the WBC (leukocyte) chamber; the engineers were tracing which valve is responsible for draining it.

What was checked

Cause

The suspected cause, reported in the case, was dirt accumulating inside the drain valve, restricting flow and degrading the pumping of fluid out of the chamber.

The cause is probable and was not fully confirmed.

What was done

The engineers found the recommended fix was to disassemble and clean the valve, since dirt builds up in it. The outcome was reported as probable: the specific valve and the final result were not confirmed in the case.

Safety

Before opening the fluid path, isolate the analyzer from mains power, and treat all internal fluids and wetted surfaces as potentially infectious biohazard; wear gloves and eye protection.

Lesson: When a chamber will not drain, trace the fluid path to the responsible valve and clean it before considering replacement.