Mindray · BC-30s

Mindray BC-30s: WBC chamber electrode looks corroded or broken

What happened

On a Mindray BC-30s hematology analyzer, the electrode in the WBC counting chamber appeared eroded flush with the chamber wall or broken off, while the electrode in the RBC chamber was long and spanned the full width of its chamber.

What was checked

Cause

The engineers found the probable cause to be chemical degradation of the WBC chamber electrode after long service: the electrode had corroded away, which explained its flush, broken-looking appearance next to the intact long electrode in the RBC chamber.

The cause is probable and was not fully confirmed.

What was done

The fix was replacement of the degraded electrode with a new one, with access and any post-replacement checks done per the manufacturer's service documentation. As reported in the case, these electrodes degrade over time, on the order of every two to three years, and on a healthy instrument the electrodes in both chambers look the same. The outcome was reported as probable, so counting performance should be verified after the swap.

Safety

Counting chambers hold residues of blood-derived fluid, so power down and isolate the analyzer before accessing the chambers, and handle all removed parts and spills as biohazardous, wearing gloves and eye protection.

Lesson: On analyzers with twin counting chambers, compare paired components against each other: an electrode that looks eaten away or shortened next to its twin is a worn consumable, not the factory condition.