Mindray · BC-3300

Mindray BC-3300: platelet background and HGB=1 on blank after reagents froze

What happened

After the reagents were frozen (reported storage range +2 to +30 °C), the analyzer showed a platelet background, and a blank cycle gave HGB=1 instead of 0, as reported in the case.

What was checked

Cause

Probable, not finally confirmed: the diluent was damaged by freezing, possibly with sediment formed, causing the platelet background and the small blank offset on HGB. The lyser and rinse were considered unlikely contributors.

The cause is probable and was not fully confirmed.

What was done

No confirmed fix was recorded in the case. The engineers' practical conclusion: after a freeze, take the diluent (and any other reagent in doubt) out of service, replace it with fresh stock kept within the reported +2 to +30 °C range, and confirm a clean blank and passing controls before returning the analyzer to use.

Safety

Handle the diluent, lyser and any sample residue as biohazardous/chemical materials with gloves and eye protection, and run a blank plus quality control before returning the analyzer to patient testing.

Lesson: A freeze excursion can silently ruin reagents and skew counts, so after any cold excursion replace the questionable reagent and prove a clean blank with passing QC before trusting results.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.