Mindray BC-5150: blocked sampling needle
What happened
The hematology analyzer aspirated poorly or not at all; the sampling needle was suspected to be blocked.
What was checked
- The engineers removed the sampling needle and flushed it with water under pressure from a syringe through a short piece of tubing.
- The engineers found a clot lodged in the needle bore; an old, dried plug did not clear with flushing alone, as reported in the case.
Cause
A blood clot had dried inside the sampling needle bore and blocked aspiration.
What was done
The engineers removed the needle, flushed it with water under pressure from a syringe through tubing, and, as reported in the case, carefully warmed the needle to soften the old clot until it broke free; the cleaned needle was refitted and aspiration recovered. A replacement needle was also being ordered, since a needle that has blocked badly is not one to trust long-term.
Safety
Treat the needle and everything flushed out of it as infectious biohazard — gloves and eye protection, the needle handled as a sharp; if it is warmed, do it away from reagents and flammables and let it cool before refitting.
Lesson: Most sampling-needle failures are dried clots that the automatic flush cannot clear — pull the needle and flush it mechanically, treat warming as a last resort, and prefer a new needle over keeping a tired one in service.