Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Mindray BC-5150
Mindray · BC-5150

Mindray BC-5150: blocked sampling needle

What happened

The hematology analyzer aspirated poorly or not at all; the sampling needle was suspected to be blocked.

What was checked

Cause

A blood clot had dried inside the sampling needle bore and blocked aspiration.

What was done

The engineers removed the needle, flushed it with water under pressure from a syringe through tubing, and, as reported in the case, carefully warmed the needle to soften the old clot until it broke free; the cleaned needle was refitted and aspiration recovered. A replacement needle was also being ordered, since a needle that has blocked badly is not one to trust long-term.

Safety

Treat the needle and everything flushed out of it as infectious biohazard — gloves and eye protection, the needle handled as a sharp; if it is warmed, do it away from reagents and flammables and let it cool before refitting.

Lesson: Most sampling-needle failures are dried clots that the automatic flush cannot clear — pull the needle and flush it mechanically, treat warming as a last resort, and prefer a new needle over keeping a tired one in service.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes