Mindray · BC-5380/Мифик 18

Mindray BC-5380: suspected lytic reagent carryover in measurements

What happened

The engineers suspected that the lytic reagent was being carried over into the measurements. In this reported case the curves on the plots looked normal, so the problem was not obvious from the graphs alone.

What was checked

Cause

In this case the suspected carryover was associated with tube 19, which the engineers found heavily clogged with contamination; a partly blocked line can retain reagent and wash it into subsequent measurements.

What was done

The engineers flushed tube 19, and the analyzer reportedly returned to normal operation. No parts were replaced in the reported case.

Safety

Treat the fluid path, samples and lytic reagents as potentially infectious and chemically aggressive: wear gloves and eye protection when opening the hydraulic lines, and confirm quality control results before returning the analyzer to use.

Lesson: When reagent carryover is suspected, inspect and flush the hydraulic tubing first — a partly blocked line can contaminate subsequent measurements even when the curves look normal.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.