Mindray · BC-5800

Mindray BC-5800: recurring WBC chamber error

What happened

On a hematology analyzer, the instrument kept flagging a chamber error during operation.

What was checked

Cause

The engineers attributed the fault to the float in the WBC chamber: a sticky, platelet-rich mass was adhering to it, so the float no longer returned freely and the chamber check kept failing.

The cause is probable and was not fully confirmed.

What was done

Mechanical cleaning: remove and disassemble the float, wash off the adhering mass and confirm it moves freely again. In this case the outcome was probable — the cleaning was the recommended next step based on similar experience, not a confirmed completed repair.

Safety

Treat all fluids, residues and removed parts from the chamber as potentially infectious biohazard: wear gloves and eye protection, and isolate the analyzer from the power supply before opening the fluid path.

Lesson: Recurring chamber errors in hematology analyzers often trace back to a float that no longer moves freely, so check it for residue before suspecting electronics or sensors.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.