Mindray · BC-DxH900

Mindray BC-DxH900: Peaks on blank, one WBC aperture erratic

What happened

The engineers reported peaks on the histogram during a blank measurement, and one of the three WBC apertures in the CBC block (as reported in the case) behaved erratically. No error codes appeared; the problem was visible on the blank graph and in the aperture voltage readings.

What was checked

Cause

The engineers found contamination of the WBC aperture: debris had deposited on the aperture, disturbing the pulses on that channel even with particle-free blank fluid. This was the suspected cause in this one case, not a general rule.

The cause is probable and was not fully confirmed.

What was done

The recommended fix was a course of flushes/washes of the affected aperture with the manufacturer-approved cleaning solution, with blank runs in between to confirm the graph returns to normal. The outcome was only probable: flushing was expected to clear the contamination, but this was not confirmed in the case. If flushes do not help, deeper cleaning of the aperture is the next step.

Safety

Treat everything leaving the fluidics path — flush waste, spent cleaning solution, removed tubing and parts — as potentially infectious biohazard: gloves and eye protection during flushes, and decontaminate any spillage immediately.

Lesson: A spiky blank on a single WBC aperture usually points to aperture contamination rather than electronics — compare channels on the blank graph first and flush early, before the debris hardens.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.