Mindray BS-200: droplets left in every cuvette after wash
What happened
After the wash cycle, residual droplets remained in the corners of every cuvette, and users suspected falsely elevated results. No error codes were reported in the case.
What was checked
- The engineers checked the cleanliness of the wash station and found dispense and aspiration consistent across all positions.
- They inspected the cuvettes immediately after a wash cycle and saw droplets persisting in the corners.
Cause
Probable weeping from the wash station after aspiration — in this case attributed to the absence of check valves in the fill and aspirate needle lines, allowing fluid to drip back into the cuvettes.
The cause is probable and was not fully confirmed.
What was done
The recommended fix (probable, not yet confirmed in this case) was to fit check valves on the fill and aspirate needle lines of the wash station so fluid stops dripping back after each cycle, in line with the manufacturer's service documentation; the engineers advised confirming the result with carry-over testing and QC before trusting sample values.
Safety
Treat the wash station and fluid path as biohazardous: decontaminate and isolate power before any disassembly, and confirm QC passes before returning the analyzer to patient samples.
Lesson: Droplets left in every cuvette after wash point to the wash station's fluid path (e.g. back-dripping) rather than individual cuvettes, so check the station hardware and verify carry-over before trusting the results.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.