Mindray BS-200: stirrer intermittently fails to run
What happened
On a Mindray BS-200 biochemistry analyzer, the stirrer would intermittently fail to run — some mixing steps were simply skipped, with no error reported.
What was checked
- Inspected the stirrer motor and its brushes for wear.
- Checked the wiring and connectors for continuity and damage.
- Examined the vertical movement shaft of the stirrer assembly and found it binding from worn-out grease.
Cause
The engineers found that the vertical movement shaft was binding intermittently because the grease had dried out; the motor and wiring checked out fine.
What was done
The engineers removed the old grease and re-greased the vertical movement shaft; after that, no missed stirrer actuations were reported, and the case was closed as solved.
Safety
Isolate the analyzer from mains power before working on the stirrer drive, treat residues around the mixing position as biohazard, and confirm smooth movement and a complete mixing cycle before returning the analyzer to use.
Lesson: Check the mechanics before swapping electrical parts — dried grease on a slide or shaft can look exactly like a motor or wiring fault.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.