Mindray BS-200: triglycerides over-read in combined test runs
What happened
When triglycerides were run together with other chemistries in one combined flow, the results came out strongly elevated; when triglycerides were run as a separate, standalone flow, the values were normal.
What was checked
- Wash quality of the instrument was checked, as reported in the case; no obvious wash fault was found.
- The engineers found the key sign: results changed with the run mode — normal in a triglyceride-only run, high in the combined run — which pointed to carryover between chemistries rather than a wash or optics problem.
Cause
The engineers found carryover contamination: when the analyzer ran triglycerides immediately after lipid tests (cholesterol, HDL, LDL, lipase) in the same reaction flow, traces left in the flow cell carried over into the triglyceride measurement and inflated the results.
The cause is probable and was not fully confirmed.
What was done
In the analyzer's settings, the test distribution was changed from the speed-optimized ('optimal time') mode to grouping by tests, so triglycerides are not scheduled straight after lipid chemistries in the same flow. The outcome in this case was reported as probable: the carryover was expected to disappear; confirm with control material and a comparison run before routine use.
Safety
Treat samples, controls and reaction mixtures as potentially infectious biohazard; follow the manufacturer's instructions and the laboratory's biosafety rules when working around the sampling and wash station, and run quality-control material before returning the analyzer to routine use.
Lesson: Chemistry results can depend on test order: carryover between neighbouring chemistries can bias results even when the wash looks fine, so check the scheduling settings before suspecting hardware.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.