Mindray · BS-200E

Mindray BS-200E: reagent rotor temperature spikes then falls back

What happened

In the reported case, a Mindray BS-200E chemistry analyzer showed unstable reagent rotor temperature: during operation it rose from the 37 °C setpoint to 39–43 °C, then dropped sharply back, with no error codes logged.

What was checked

Cause

Probable cause, based on a service assessment shared in the case: the P5 aspiration pump was not emptying the cuvettes properly. Residual liquid and condensate in the rotor area disturbed the thermal balance, so the heater control pushed the temperature above setpoint and then pulled it back — rather than a failing temperature sensor, as first suspected.

The cause is probable and was not fully confirmed.

What was done

The suggested fix was to restore proper cuvette emptying: service or replace the P5 aspiration pump, then confirm over several full runs that the rotor holds its 37 °C setpoint. Because the cause was only probable, the engineers treated the pump replacement as a verification step — if the spikes persist, look at condensate paths and the temperature sensor next.

Safety

Treat cuvette residue and condensate as potentially infectious, wear gloves and eye protection, and before returning the analyzer to use confirm temperature stability and perform an electrical safety test, since liquid sits close to powered components.

Lesson: When a thermostat overshoots and then recovers while fans are clean and no errors are logged, suspect residual liquid or condensate disturbing the thermal control before blaming the temperature sensor.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.