Mindray · BS-240 Pro

Mindray BS-240 Pro: summer condensation on the reagent disk

What happened

On one Mindray BS-240 Pro, condensation kept forming on the reagent disk, but only during the summer months; no error codes were reported in the case.

What was checked

Cause

The cause was not confirmed in this case, but the pattern points to warm, humid summer air condensing on the rotor surface, which stays cooler; an open or frequently opened cover makes it worse.

The cause is probable and was not fully confirmed.

What was done

The recommendation in this case was to lower the room temperature and dry the air with air conditioning, and to place silica gel in a cloth pouch inside the rotor, positioned so it cannot touch the sampling probe. The outcome was reported as probable: similar reports are common for this model, but the final solution was not confirmed.

Safety

Condensation means moisture near electronics and optics: isolate the analyzer from the mains before placing a desiccant pouch or wiping the rotor, secure the pouch so it cannot touch the moving disk or the probe, and confirm everything is fully dry before returning the unit to use.

Lesson: Recurring seasonal condensation usually starts in the room climate, so sort out temperature and humidity first before suspecting the analyzer itself.