Mindray BS-2800: Liquid ejected back during aspiration
What happened
During aspiration the analyzer pushes liquid back out instead of drawing the sample in cleanly. No error codes were reported in the case.
What was checked
- The engineers suggested checking the probe-side valve for a fault or clog.
- They inspected the probe washer for salt deposits and dried blood residue.
- They also checked the plastic bushing of the washer for wear.
Cause
The engineers found the probe washer contaminated with salts and blood residue; wear of its plastic bushing was considered a possible additional cause. The diagnosis was probable, based on the discussion in this single case.
The cause is probable and was not fully confirmed.
What was done
The recommended fix was to clean the probe washer thoroughly, and replace the washer assembly if the plastic bushing is worn. The outcome was probable — the engineers expected the cleaning to stop the backflow, but it was not confirmed in this case.
Safety
Washer residue containing blood traces is potentially infectious — wear gloves and eye protection during cleaning and handle all fluids as biohazardous until rinsed away.
Lesson: When an analyzer ejects liquid back at aspiration, start with the simple mechanical suspects — a dirty probe washer or a worn bushing — before moving on to valves or electronics.