Mindray BS-2800: Liquid ejected back during aspiration

What happened

During aspiration the analyzer pushes liquid back out instead of drawing the sample in cleanly. No error codes were reported in the case.

What was checked

Cause

The engineers found the probe washer contaminated with salts and blood residue; wear of its plastic bushing was considered a possible additional cause. The diagnosis was probable, based on the discussion in this single case.

The cause is probable and was not fully confirmed.

What was done

The recommended fix was to clean the probe washer thoroughly, and replace the washer assembly if the plastic bushing is worn. The outcome was probable — the engineers expected the cleaning to stop the backflow, but it was not confirmed in this case.

Safety

Washer residue containing blood traces is potentially infectious — wear gloves and eye protection during cleaning and handle all fluids as biohazardous until rinsed away.

Lesson: When an analyzer ejects liquid back at aspiration, start with the simple mechanical suspects — a dirty probe washer or a worn bushing — before moving on to valves or electronics.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.