Mindray · BS-3000

Mindray BS-3000: corroded electrode plates in WBC/RBC counting chambers

What happened

The electrode plates inside the WBC/RBC counting chambers were corroded through and needed replacement.

What was checked

Cause

Corrosion of the original electrode plates in the WBC/RBC chambers — constant contact with the reagent environment gradually eats them away.

What was done

The engineers replaced the corroded plates with nickel strips sold for battery spot welding: the thickness matches the originals and the material solders easily. In the reported case such a replacement held up for about two years. This is a field substitute from a single case, not the manufacturer's part or procedure; the resistance of the substitute raised doubts in the discussion, so cell counts should be confirmed with control samples before the analyzer returns to routine work.

Safety

Chamber parts and electrodes are contaminated with blood-derived material — wear gloves and eye protection during disassembly, disinfect all removed parts before disposal, and verify counting performance with control samples before the analyzer is returned to use.

Lesson: Electrode plates in impedance counting chambers are wearing parts that corrode over time, and any substitute material must be checked for resistance and verified with control samples before the analyzer goes back into service.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.