Mindray BS-3000: corroded electrode plates in WBC/RBC counting chambers
What happened
The electrode plates inside the WBC/RBC counting chambers were corroded through and needed replacement.
What was checked
- Inspected the electrode plates in the chambers and confirmed heavy corrosion damage.
- Evaluated the substitute material by its electrical resistance and compared it with nickel strips used for battery spot welding.
Cause
Corrosion of the original electrode plates in the WBC/RBC chambers — constant contact with the reagent environment gradually eats them away.
What was done
The engineers replaced the corroded plates with nickel strips sold for battery spot welding: the thickness matches the originals and the material solders easily. In the reported case such a replacement held up for about two years. This is a field substitute from a single case, not the manufacturer's part or procedure; the resistance of the substitute raised doubts in the discussion, so cell counts should be confirmed with control samples before the analyzer returns to routine work.
Safety
Chamber parts and electrodes are contaminated with blood-derived material — wear gloves and eye protection during disassembly, disinfect all removed parts before disposal, and verify counting performance with control samples before the analyzer is returned to use.
Lesson: Electrode plates in impedance counting chambers are wearing parts that corrode over time, and any substitute material must be checked for resistance and verified with control samples before the analyzer goes back into service.