Mindray · BS-380

Mindray BS-380: sample probe not detecting liquid level on the reaction disc

What happened

Reported in the case: during routine operation the sample probe of a Mindray BS-380 stopped recognising the liquid level over the reaction disc (RA1 position), the software flagged a level-detection failure, and suspicion fell on the serum needle assembly.

What was checked

Cause

Not finally confirmed. The working theory in this case: the level-sensing calibration of the serum probe had drifted — after contamination, cleaning cycles or component ageing — so the probe stopped recognising the meniscus over the reaction disc; a genuine sensor or probe defect was considered less likely.

The cause is probable and was not fully confirmed.

What was done

The recommended fix was an auto-calibration of the needle level sensor — a procedure from the manufacturer's service documentation, done by authorised personnel. The outcome was reported as probable: the error was expected to clear after calibration, but long-term confirmation was not given.

Safety

Treat the probe and everything it has touched as biohazardous and wear gloves, disconnect the analyzer from the mains and let moving parts stop before working around the needle or its sensor board, and verify level detection with control runs before the analyzer returns to sample processing.

Lesson: A 'needle failed' level-detection error on a chemistry analyser is often just a drifted sensor calibration — check the probe hardware first, then restore calibration through the proper service procedure instead of replacing parts.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.