Mindray · DC-40 (Mylab 40?)

Mindray DC-40: keyboard dead except probe selection

What happened

On this ultrasound unit the keyboard stopped responding to everything except probe selection. After a BIOS reflash the machine still would not let anyone into the menus needed to install the software.

What was checked

Cause

The reported diagnosis was a failed motherboard, described in the case as a common failure on this model. This is one case, not a rule.

The cause is probable and was not fully confirmed.

What was done

A motherboard replacement was recommended, quoted in the case at roughly €3000; the fix was probable, not confirmed, and with a repair cost like that, writing the machine off was on the table.

Safety

Before any work on the main board, disconnect the machine from the mains, wait for the internal supplies to discharge, work with ESD protection, and run the full electrical safety test before the unit returns to clinical use.

Lesson: A keyboard that answers only the probe selector usually points to the main board, and a BIOS reflash that changes nothing means it is time to stop chasing software.