Mindray DC-N6: no power-on
What happened
The ultrasound does not start at all: no boot, the keyboard stays dark, and (reported in the case) a fourth indicator lights briefly and then goes out.
What was checked
- Confirm the symptom pattern: fully dead unit, dark keyboard, and the indicator that lights and then drops out (reported in the case).
- The engineers advised checking the DC-DC board mounted above the AC-DC unit: on a healthy board three status LEDs should be lit (reported in the case). In this case all of them stayed dark.
- Before condemning the board, verify the mains cord, inlet and connections to the board, since a dead DC-DC can also mean nothing is feeding it.
Cause
Probable failure of the DC-DC board: with its three status LEDs dark, the engineers concluded the board is likely not producing the internal supply voltages the system needs to start. The diagnosis remained probable, not confirmed by bench test.
The cause is probable and was not fully confirmed.
What was done
No repair was completed in the case — the discussion ended at diagnosis. The likely next step is repair or replacement of the DC-DC board; any board-level work on this power path follows the manufacturer's service documentation and is done by authorised personnel, followed by a full function and electrical safety check before return to use.
Safety
Disconnect the unit from the mains and allow the power-supply capacitors to discharge before opening the AC-DC/DC-DC section — switch-mode supplies hold dangerous stored charge.
Lesson: A completely dead system with a status-LED bank dark on the internal DC-DC board is a fast first clue — check those LEDs before swapping major assemblies.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.