Mindray Eliys Duo: no buffer supply on the yellow channel
What happened
Buffer from the external container was not delivered on the yellow channel; the channel pump did not run, although fluid moved when the line was flushed manually with a syringe.
What was checked
- The engineers flushed the fluid line with a syringe — liquid moved, so the tubing and the path from the external container were not blocked.
- They measured supply voltage at the pump during a delivery command — voltage was present, but the pump still did not pump.
Cause
The engineers found that the pump itself had failed: with supply voltage present it would not pump. It is a sealed, non-serviceable unit, so it could not be rebuilt on site and had to be replaced.
What was done
The pump was replaced; the case was reported as solved. As a temporary measure in this case, the pump from an unused channel was moved to the yellow channel, since the channel pumps are identical, and buffer delivery was restored.
Safety
Isolate the analyzer from mains power before removing or fitting the pump, and treat all fluid lines, buffers and any spills as potentially infectious material — work in gloves and decontaminate immediately.
Lesson: When a pump has supply voltage but does not pump while fluid flows freely on a manual syringe flush, suspect the pump itself — and on multi-channel analyzers an identical pump from an unused channel can quickly confirm the fault or bridge the gap until the replacement arrives.