Mindray Resona / DC-X6: system hangs at the boot logo for up to an hour
What happened
After power-on the system sits on the logo screen: boot either does not move or takes very long — reported in the case as 15–30 minutes, sometimes up to an hour — after which the system may still load normally.
What was checked
- Do not cut power while the logo is on the screen — in this case the system eventually finished booting after a long wait.
- After a full boot, photograph the page with the installed options before touching anything.
- Boot progress was watched on the console startup output, accessed as described in the manufacturer's service documentation; the boot was seen to stall around filesystem activity.
Cause
In this single case the engineers found the probable cause was filesystem errors on the system hard drive, with the drive itself likely nearing end of life; each boot re-reads damaged areas and crawls.
The cause is probable and was not fully confirmed.
What was done
The engineers let the system boot fully without power-cycling it, recorded the options data, then prepared to swap or replace the system hard drive. Viewing the startup messages and confirming the filesystem check was a procedure from the manufacturer's service documentation, done by authorised personnel. Full recovery after the replacement was probable, not confirmed in the case.
Safety
Disconnect the scanner from mains and let internal capacitors discharge before opening the console, handle the drive with ESD precautions, back up the options data first, and run an electrical safety test before returning the unit to use.
Lesson: Long hangs at the boot logo usually point to failing storage: let the unit finish booting, save the options data, then replace the drive before it fails completely.