Mindray · Super GL

Mindray Super GL: sample error on first calibration

What happened

The analyzer aspirates samples normally, but the first calibration stops with a sample error. The fluid system checks out and the buffer was fresh.

What was checked

Cause

Someone had earlier replaced the needle-to-sensor tube with a wider, longer one. The extra volume in the sample path changed the sample volume reaching the sensor, so calibration flagged a sample error and stopped the analyzer.

The cause is probable and was not fully confirmed.

What was done

The engineers cut the non-standard tube down to the stock length, and calibration then passed. The outcome is only probable: reported readings for normal and pathology levels came out almost twice as low, which means the wider bore still distorts the sample volume. The stock needle-to-sensor tube (0.89 mm × 12.5 cm, per the case) will most likely need to be installed, followed by calibration and quality-control checks.

Safety

Treat the sample path and any removed tubing as biohazardous: wear gloves, disinfect spills, and do not return the analyzer to testing until calibration and quality control pass.

Lesson: Non-standard tubing can pass daily sampling yet break volume-dependent calibration — compare internal tube dimensions with the stock specification before swapping other parts.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.