Mindray Super GL: sample error on first calibration
What happened
The analyzer aspirates samples normally, but the first calibration stops with a sample error. The fluid system checks out and the buffer was fresh.
What was checked
- The engineers checked the fluid system and the fresh buffer first — nothing wrong there.
- A suspect valve had been replaced earlier in this case, but the sample error at calibration stayed.
- The sensor and the sample path were then checked; the tubing from the needle to the sensor turned out non-standard — reported in the case as 1.14 mm × 17 cm instead of the stock 0.89 mm × 12.5 cm.
Cause
Someone had earlier replaced the needle-to-sensor tube with a wider, longer one. The extra volume in the sample path changed the sample volume reaching the sensor, so calibration flagged a sample error and stopped the analyzer.
The cause is probable and was not fully confirmed.
What was done
The engineers cut the non-standard tube down to the stock length, and calibration then passed. The outcome is only probable: reported readings for normal and pathology levels came out almost twice as low, which means the wider bore still distorts the sample volume. The stock needle-to-sensor tube (0.89 mm × 12.5 cm, per the case) will most likely need to be installed, followed by calibration and quality-control checks.
Safety
Treat the sample path and any removed tubing as biohazardous: wear gloves, disinfect spills, and do not return the analyzer to testing until calibration and quality control pass.
Lesson: Non-standard tubing can pass daily sampling yet break volume-dependent calibration — compare internal tube dimensions with the stock specification before swapping other parts.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.