Mindray · WATO EX-35

Mindray WATO EX-35: bellows rises crooked or only halfway

What happened

The ventilator bellows rose at an angle or only to half height and then seemed to stick. After cleaning it behaved normally for about two days, as reported in the case, then the problem returned. No error codes were reported.

What was checked

Cause

In this case the engineers linked the symptom to a passive scavenging line connected to a forced room extraction without compensating openings: the resulting negative pressure pressed on the bellows, so it rose unevenly or only partway and appeared to stick.

The cause is probable and was not fully confirmed.

What was done

The probable fix was to relieve the excess vacuum by adding compensating openings in the scavenge hose, followed by a repeat leak check. Any change to the scavenging path must follow the manufacturer's service documentation and be carried out by authorised personnel.

Safety

Before the machine returns to use, run a full leak and ventilation check with normal bellows travel confirmed, since any scavenging change also affects capture of waste anaesthetic gases in the room.

Lesson: An ascending bellows that misbehaves on and off is not always dirty — check how the scavenging system is connected before rebuilding the bellows.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.