Mindray WATO EX-35: circuit fault stays after circuits and bellows replacement
What happened
The machine reported a breathing circuit fault. Fresh circuits and a new bellows did not clear it.
What was checked
- Breathing circuits were replaced; the fault stayed.
- Bellows were replaced with new ones; the fault stayed.
- Photos of the valve area were requested to check whether sorbent dust had got under the valve — a pattern known from training.
Cause
Probable cause in this case: sorbent dust got under the valve and kept the circuit fault active.
The cause is probable and was not fully confirmed.
What was done
Disassemble and clean the valve assembly; in this case that was reported as the probable fix. In a similar case the same symptom was solved by replacing the pressure gauge — if cleaning does not help, check the gauge next.
Safety
After any work on the breathing circuit or valves, run the full leak and ventilation checks from the manufacturer's manual before the machine goes back to clinical use — a sealing fault here directly affects the patient's ventilation.
Lesson: When fresh circuits and bellows do not clear a circuit fault, stop swapping consumables and inspect the valve and pressure gauge for sorbent dust.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.