Mindray · WATO EX-35

Mindray WATO EX-35: circuit fault stays after circuits and bellows replacement

What happened

The machine reported a breathing circuit fault. Fresh circuits and a new bellows did not clear it.

What was checked

Cause

Probable cause in this case: sorbent dust got under the valve and kept the circuit fault active.

The cause is probable and was not fully confirmed.

What was done

Disassemble and clean the valve assembly; in this case that was reported as the probable fix. In a similar case the same symptom was solved by replacing the pressure gauge — if cleaning does not help, check the gauge next.

Safety

After any work on the breathing circuit or valves, run the full leak and ventilation checks from the manufacturer's manual before the machine goes back to clinical use — a sealing fault here directly affects the patient's ventilation.

Lesson: When fresh circuits and bellows do not clear a circuit fault, stop swapping consumables and inspect the valve and pressure gauge for sorbent dust.