Mindray WATO EX-35: delivered volume lower than set
What happened
The machine delivers less volume than set: the on-screen graphs show one figure, while the volume observed on the bellows is clearly smaller. No error codes were reported in the case.
What was checked
- Run the routine flow sensor calibration, then re-check the delivered volume.
- Verify the delivered volume with a calibrated flow analyzer and compare it with the set and displayed values.
- Leak-test the breathing circuit and inspect the bellows, tubing, valves and connectors for wear or poor seating.
Cause
No confirmed cause. The working suspicion of the engineers was reading drift of the flow sensors; a leak or wear in the breathing circuit had to be ruled out first, as it can give the same picture.
The cause is probable and was not fully confirmed.
What was done
The recommended path in the case: start with the routine flow sensor calibration and re-check the delivered volume. If a flow analyzer confirms the mismatch, proceed to the internal flow sensor calibration — a procedure from the manufacturer's service documentation, done by authorised personnel, with a flow analyzer and the matching adapter. The outcome was probable; the final result was not confirmed in the case.
Safety
Do not return the machine to a patient until volume accuracy is confirmed with a calibrated flow analyzer and the breathing circuit passes the leak test and acceptance checks per the manufacturer's documentation.
Lesson: When the displayed and delivered volumes disagree, confirm with an independent flow analyzer and rule out the breathing circuit before blaming or recalibrating the sensors.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.