Mindray Wato EX-65: water ingress cracked the blower housing
What happened
The engineers found a crack in the blower (volute) housing with water streaks — water from the compressor had entered the machine's fan.
What was checked
- Inspect the mounting plate and surrounding area for water streaks.
- Check the internal flow sensor for moisture and deposits.
- Check the PEEP valve and the safety valve for water and corrosion.
- Trace the tubing towards the Monitoring Board to see whether water reached the sensors.
- Assess the amount of oxidation on the valves — this decides the scope of replacement.
Cause
In this case, water carried over from the compressor entered the blower; with drive gas set to air, the accumulated water was enough to crack the housing from the inside.
The cause is probable and was not fully confirmed.
What was done
As discussed in the case: if oxidation is minimal, replacing only the blower (volute) may be enough; where valves show oxidation, the flow valve and possibly the PEEP and safety valves need replacement — on some other platforms, cleaning plus a single flow valve has been sufficient in similar cases. The outcome was considered probable, not confirmed.
Safety
Disconnect the machine from the mains before opening the pneumatic compartment, treat the wetted gas path as contaminated, and complete electrical safety and leak tests before the ventilator is returned to use.
Lesson: Water from a compressor or supply line can travel far beyond the point of entry — after any ingress, trace the whole pneumatic path, including the flow sensor, valves and boards, not just the obviously damaged part.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.