Monnal T75: expiratory valve error on self-test
What happened
The ventilator flags an expiratory valve fault when it runs its self-test.
What was checked
- Continuity-tested all the valves; every one read as serviceable.
- Checked the driver transistors on the valve board; no failures found.
- Pulled and inspected the black rubber seat ring under the expiratory valve.
Cause
The engineers found the black rubber O-ring that the expiratory valve seats against cracked around its edges, most likely from aging. In this one case the broken seal at the seat was enough for the self-test to blame the valve; the cause is reported as probable.
The cause is probable and was not fully confirmed.
What was done
The ring lifts out on its own, so it was replaced by itself and the valve kept. Had the ring been intact, the valve itself would have been the next part to swap. After either replacement the ventilator needs recalibration against a gas flow analyzer per the service documentation. The outcome in this case was reported as probable, not confirmed under full load.
Safety
Before the ventilator returns to a patient, confirm the repair with a leak and calibration check on a gas flow analyzer, and disinfect every expiratory-path part you handled, since patient gas can leave bioburden there.
Lesson: An expiratory valve fault at self-test can hide in the cheapest seal on the machine — inspect the seat ring before ordering electronics or a valve assembly.
Covered in the course: Ventilator Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.