MTL Integris 18.0.3: DICOM CD burn errors on known-good discs

What happened

Every attempt to burn a DICOM CD ended in an error. A clean DVD-R was in the drive (CD-R was also tried), the same discs burned and read normally from Windows, and replacing the drive changed nothing.

What was checked

Cause

The engineers found that the DICOM CD settings contained a link to a nonexistent file; the recording routine stumbled on this dead reference before it ever reached the drive.

What was done

The fix was simply deleting the stale reference from the settings/dicomcd section; after that DICOM CD burning worked again with the original drive.

Safety

Before pulling or swapping the optical drive, shut the workstation down and disconnect it from the mains, follow ESD precautions, and carry out an electrical safety check before the station returns to clinical use.

Lesson: When burning fails on good media with a good drive, move to the software side early — a stale file reference in the DICOM CD settings can block recording as effectively as dead hardware.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.