MTL Integris 18.0.3: DICOM CD burn errors on known-good discs
What happened
Every attempt to burn a DICOM CD ended in an error. A clean DVD-R was in the drive (CD-R was also tried), the same discs burned and read normally from Windows, and replacing the drive changed nothing.
What was checked
- Checked the media itself: burned and read discs from under Windows to rule out bad or incompatible blanks.
- Swapped the optical drive — the fault stayed with the station, so the drive was not the culprit.
- With hardware ruled out, went through the DICOM CD section of the settings and found a reference to a file that no longer existed on the workstation.
Cause
The engineers found that the DICOM CD settings contained a link to a nonexistent file; the recording routine stumbled on this dead reference before it ever reached the drive.
What was done
The fix was simply deleting the stale reference from the settings/dicomcd section; after that DICOM CD burning worked again with the original drive.
Safety
Before pulling or swapping the optical drive, shut the workstation down and disconnect it from the mains, follow ESD precautions, and carry out an electrical safety check before the station returns to clinical use.
Lesson: When burning fails on good media with a good drive, move to the software side early — a stale file reference in the DICOM CD settings can block recording as effectively as dead hardware.