Neusoft NeuViz 16: workstation data drive nearly full, RAID risk before disk swap

What happened

On the CT workstation, the local D: drive holding the patient database — reported in the case as 192 GB — was almost full, while E:, reported as 900 GB and holding raw scan data, was nearly empty. The idea was to swap the disk contents and reassign drive letters in Disk Management.

What was checked

Cause

The engineers found no fault in the array itself: in this one case the database volume had simply filled up with older patient data, while the much larger raw-data volume sat nearly unused.

What was done

The disk swap and letter reassignment were ruled out: with the disks combined in a RAID array, such manipulations would destroy the array. The recommendation was to leave the disks as they are, verify the array layout in the BIOS, and simply free space on the full data drive by clearing old data in line with the site's data-retention rules. The issue was solved in this case.

Safety

Before opening the workstation, shut it down and disconnect it from the mains, and after any work inside the enclosure run the electrical safety test before the unit returns to use.

Lesson: Before swapping disks or drive letters on an integrated system, confirm whether the volumes sit on a RAID array — freeing or archiving old data is usually the safer first move.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.