Newport · HT70

Newport HT70: Intermittent drift of delivered volume and respiratory rate

What happened

As reported in the case: the ventilator ran normally for about two days, then the delivered volume started drifting and the respiratory rate reading jumped. The fault appeared rarely, with no error codes, and the unit behaved normally in between episodes.

What was checked

Cause

The engineers suspected water accumulating in the breathing circuit: in this case it could make the machine auto-cycle and measure volume and rate incorrectly. This remained a probable cause — it was not confirmed during the case, and no hardware fault was found.

The cause is probable and was not fully confirmed.

What was done

A confirmed fix was not reported. The practical step agreed in the case was to eliminate the water: drain the circuit, check the water traps and humidifier, then re-verify volume and rate readings. If the fault returned, the recommendation was to capture episodes on photo/video and send the unit for manufacturer-level paid diagnostics. No parts were replaced.

Safety

Treat condensate from the breathing circuit as potentially infectious: drain it away from the patient side wearing gloves and eye protection, and complete a full performance verification with a test lung before the ventilator returns to clinical use.

Lesson: An intermittent fault with no error codes is captured, not guessed — document the episode first and rule out simple physical causes like water in the circuit before blaming electronics.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.