Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Nihon Kohden Celltac
Nihon Kohden · Celltac

Nihon Kohden Celltac: error at power-on after battery replacement

What happened

After the backup battery was replaced, the analyzer reported an error at every power-on and did not continue into normal operation.

What was checked

Cause

In this case the error was most likely caused by the loss of stored panel calibration data when the battery was disconnected, so the unit required recalibration at the next start.

What was done

The fix was the panel calibration routine at power-on — a procedure from the manufacturer's service documentation, done by authorised personnel. After calibration was completed, the analyzer powered up without the error.

Safety

Before opening the analyzer to change the battery, disconnect it from the mains and let it power down fully; handle the cell with care — do not short it, observe polarity when fitting the new one, and dispose of the old cell according to local rules.

Lesson: A backup-battery change can wipe stored calibration data, so check the service documentation for post-battery procedures before returning the analyzer to use.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes