Nihon Kohden Celltac: error at power-on after battery replacement
What happened
After the backup battery was replaced, the analyzer reported an error at every power-on and did not continue into normal operation.
What was checked
- The engineers replaced the backup battery as the first step; the error still appeared at the next power-on.
- They then ran the calibration routine available at power-on; the calibration menu opened, and after calibration was completed the analyzer started normally.
Cause
In this case the error was most likely caused by the loss of stored panel calibration data when the battery was disconnected, so the unit required recalibration at the next start.
What was done
The fix was the panel calibration routine at power-on — a procedure from the manufacturer's service documentation, done by authorised personnel. After calibration was completed, the analyzer powered up without the error.
Safety
Before opening the analyzer to change the battery, disconnect it from the mains and let it power down fully; handle the cell with care — do not short it, observe polarity when fitting the new one, and dispose of the old cell according to local rules.
Lesson: A backup-battery change can wipe stored calibration data, so check the service documentation for post-battery procedures before returning the analyzer to use.