Academy / Field fault notes / Nihon Kohden MicroCC-20
Nihon Kohden · MicroCC-20

Nihon Kohden MicroCC-20: PLT counts read low during calibration

What happened

As reported in the case, PLT values came out low — roughly 75% of the lower expected level — calibration would not let the value go any lower, and a blank run showed background around 5.

What was checked

Cause

The engineers considered the most likely cause to be a calibrator whose assigned PLT values did not match its stated specifications; interference from the UPS remained a possible contributor. The conclusion was probable, not confirmed.

The cause is probable and was not fully confirmed.

What was done

Calibration was redone against fresh human blood, and the sample-type calibration coefficients were corrected. The coefficient adjustment was a procedure from the manufacturer's service documentation, done by authorised personnel. The outcome was reported as probable.

Safety

Fresh blood and the analyzer's fluid path are potentially infectious: work with gloves, contain any spills, and decontaminate the sampling area after calibration.

Lesson: When the counter will not reach the calibration target, suspect the calibrator's assigned values and mains noise before blaming the counting channel.